套裝砝碼使用校正與維護 套裝砝碼介紹 組合形式 1mg-500mg——mg:1,2,2*,5,10,20,20*,50,100,200,200*,500 1mg-100g——mg:1,2,2*,5,10,20,20*,50,100,200,200*,500 g:1,2,2*,5,10,20,20*,50,100 1mg-200g——mg:1,2,2*,5,10,20,20*,50,100,200,200*,500 g:1,2,2*,5,10,20,20*,50,100,200,200* 1mg-500g——mg:1,2,2*,5,10,20,20*,50,100,200,200*,500 g:1,2,2*,5,10,20,20*,50,100,200,200*,500 1mg-1kg——mg:1,2,2*,5,10,20,20*,50,100,200,200*,500 g:1,2,2*,5,10,20,20*,50,100,200,200*,500,1000 1mg-2kg——mg:1,2,2*,5,10,20,20*,50,100,200,200*,500 g:1,2,2*,5,10,20,20*,50,100,200,200*,500,1000,2000,2000* 1mg-5kg——mg:1,2,2*,5,10,20,20*,50,100,200,200*,500 g:1,2,2*,5,10,20,20*,50,100,200,200*,500,1000,2000,2000*,5000 1mg-10kg——mg:1,2,2*,5,10,20,20*,50,100,200,200*,500 g:1,2,2*,5,10,20,20*,50,100,200,200*,500,1000,2000,2000*,5000,10000 1mg-20kg——mg:1,2,2*,5,10,20,20*,50,100,200,200*,500 g:1,2,2*,5,10,20,20*,50,100,200,200*,500,1000,2000,2000*,5000,10000,20000 套裝砝碼使用校正與維護
一、使用原則 使用時,不能用手捏,只能用鑷子夾或佩戴手套拿取; 嚴禁接觸腐蝕性物質(zhì). 嚴防碰撞或墜落,避免碰傷或劃傷. 在使用砝碼時,要"先大后小". 二、砝碼的管理 分析室使用的砝碼包括基準砝碼和日常使用砝碼(相同規(guī)格各一套)。 基準砝碼和日常使用砝碼均由有資質(zhì)的單位檢定(校準),作為公司基準器具使用。 日常使用砝碼是指在日常工作中,用來檢查天平是否正常的砝碼。 三、砝碼的校正 砝碼委托有校正資質(zhì)的單位進行定期檢定(校準) 并在檢定有效期限內(nèi)實施下一次的檢定(校正)。 砝碼的校準有效期限按照檢定報告的規(guī)定(國家規(guī)定)執(zhí)行。 四、砝碼的管理 砝碼須保存于溫度、濕度變化較小,且沒有振動的密閉容器中。 五、砝碼的使用和維護 日常使用砝碼用來實施天平的日常檢查(校正)。如在日常檢查中 出現(xiàn)偏差(結(jié)果出判定基準),則使用基準砝碼進行復(fù)檢(確認),并進行以下處置: 復(fù)檢結(jié)果仍出范圍,則判斷天平出現(xiàn)異常,按照儀器故障處理。 復(fù)檢結(jié)果正常,則判斷日常使用砝碼受到污損,須重新送檢(校準)。 砝碼不可直接用手接觸,必須使用鑷子、手套等工具,且此工具不得用于其他用途,以防止污染。 如砝碼被污染或跌落,需重新校準。 套裝砝碼使用校正與維護 |